Mike Williams, Zebulun Krahn and Curtis A. Meyer
26 July 2002
The Thomas Jefferson National Accelerator Facility (JLAB) in Newport News, VA is a new facility that is utilizing high energy electron and photon beams to study the structure of nuclear matter. JLAB is currently planning to double the energy of its accelerator with the main physics emphasis on experiments to try and explain why the constituents which build the protons and neutrons (quarks) are forever confined inside their parent particle. This particular question has been listed in the New York Times as one of the most important scientific questions of the new millennium. To attack this problem, an international group of physicists has come together to build a entirely new beam line and detector at the lab, known as Hall D. Carnegie Mellon is currently one of the leading institutions on this $35,000,000 detector. The current plan is to be able to start taking data with this detector in early '2009.
The goal of this study is to examine signals generated in a strawtube
chamber that is similar in tube-design to that planned in the GlueX
experiment. The chamber is part of the former EVA experiment at
Brookhaven and has been given to CMU. Signals were generated using
a highly collimated
source, and coincidences were recorded
on the four layers of the existing chamber. The signals were studied
using two different gas mixtures, and several different voltage
settings on eh chamber. A good understanding of the dynamic range of
the signals is needed to properly design the electronics to be used
on the chamber.
The EVA chamber consists of 2m long straw tubes each with diameter of 2.088cm. The cathode wire is a resistive wire that is nominally read out at both ends, and through charge division, the position in 'z' can be determined. (The GlueX experiment plans to accomplish this using stereolayers). The chamber has four layers of tubes, with 16 tubes in each layer. The electronics is blocked into 8 groups of tubes, with two tubes from each layer in each group (see Figure 1).
To determine the dynamic ranges of the measurements involved,
we utilized one block of eight tubes. We then concentrated
on one set of four tubes which are read out by a pre-amplifier
card. A cross-sectional diagram of a set of eight tubes is given in
Figure 1. We selected four adjacent tubes, linearly aligned,
as our detector, in this case we used tubes 2,4,6 and 8.
We read the signals from the chamber through the pre-amplifier
and record them with a Tektronix 4-channel digital oscilloscope.
A collimated
source,
, placed at a crossing angle
of 90 degrees was used to keep the flux of particles through the
chamber fairly consistent while the experiment was running.
![]() |
Data was taken by triggering off of the signal in tube 8 and recording
coincident signals in all 4 tubes. These recorded
signals were then analyzed for signal threshold, maximum
peak height (with the threshold subtracted out), pulse length, and
the integral value of the pulse (also with threshold subtracted out) using
a C++ program written for the analysis.
Data was acquired under several
sets of conditions. First, the experiment was run with a
90/10 Argon-Carbon-dioxide (
) gas mixture and the voltage
on the tubes was varied from 1700V to 1850V in 50V increments.
Then, the composition of the gases in the chamber was changed to a
50/50 Argon-Ethane (
) mixture. Data was taken at 2050V and 2100V.
![]() |
The 1800V data with the
gas mixture was also
subjected to a tube by tube analysis. This analysis allowed for
further checks of consistency and reproducibility. While the
threshold varied from tube to tube, once these threshold values were
subtracted out, the data were consistent, which allowed us to
combine the data from all four tubes. The threshold of each tube
is independent of the high voltage settings.
The maximum height of each pulse with the threshold value
subtracted out was binned into a histogram. As shown be by Figure
5, the signal peak size increases with increasing voltage, as expected.
Histograms of the peak minus threshold value for tested voltages of the
90/10
gas mixture are in Figure 6.
Histograms for the
values are in Figure 7. From these
distributions and the signal quality, it was determined that the optimal
voltage for operation with the 90/10
gas mixture
was approximately 1800V. It is clear to see that when the voltage was
increased beyond to 1850V the distribution broadened drastically and noise
in the signals also increased.
The optimal running voltage for the 50/50
gas mixture
was higher than the
mixture at 2050V. Signals here were
clean and the distributions reasonable. Data cited in the remainder
of our report will be taken from experimental trials operating under
these voltages.
![]() |
The pulse length varied greatly depending upon the gas mixture used.
The pulse length is defined by our anylsis as the duration the signal
is above 20 percent of its maximum value. The analysis does not
require that the bins included be connected. The pulse length measured
for the
is more consistent with estimated drift times when
the geometry of the tubes is taken into account. In this case, almost
all of the recorded signals yielded a length that fell at or below 200ns.
Whereas the distribution of the pulse lengths of the
mixture seems
to be broader, with many pulses extending beyond the 300ns mark.
Figure 8 shows histograms of the pulse length values for the
two gas mixtures.
![]() |
Since we are trying to define the dynamic range of the
integral value a Riemann summing technique is used to
calculate the integral values.
Histograms of integral values (peak minus threshold) for
both gas mixtures at their optimal operating voltages are shown
in Figure 9. The distribution for the
and is
more sharply peaked than the distribution for the
mixture.
The peaks appear below the 10 V-ns mark in both cases,
and the distribution of the signals extends out to 30V-ns.
Note the integral values of the
are suppressed due to
shorter pulse length in comparison to the
values.
![]() |