A Study of leakage in Partial Wave Analysis for the
HALL D Detector at Jefferson Lab
Ben Zaroukian and Jeffrey Kaditz
December 7, 2001
Abstract:
The partial wave analysis of the reactions

for two different final states have been used to detector design issues
related to resolution, low energy thresholds and particle reconstruction.
These studies indicate that the resolution of the Barrell Calorimeter is
one of the most sensitive parameters to minimizing feedthrough into the
exotic

channel.
Introduction
The Thomas Jefferson National Accelerator Facility (JLAB) in Newport News,
VA is a new facility that is utilizing high energy electron and photon beams
to study the structure of nuclear matter. JLAB is currently planning to
double the energy of its accelerator with the main physics emphasis on
experiments to try and explain why the constituents which build the
protons and neutrons (quarks) are forever confined inside their parent
particle. This particular question has been listed in the New York Times
as one of the most important scientific questions of the new millennium.
To attack this problem, an international group of physicists has come
together to build a entirely new beam line and detector at the lab, known
as Hall D. Carnegie Mellon is currently one of the leading institutions
on this $35,000,000 detector. The current plan is to be able to start
taking data with this detector in early '2007.
One of the signatures of this physics is a new type of subatomic particle
that has so called exotic quantum numbers. In order to identify when
these new particles have been created, and to measure their quantum numbers,
one performs a type of analysis called a Partial Wave Analysis. For this
sort of analysis to work, it is necessary that the experiment be designed with
it specifically in mind. As such, it is extremely important to optimize
the detector for such an analysis.
In order to optimize the Hall D detector for the physics goals, it needs to
be optimized in performing a partial wave analysis. The goal of this research
project was to study detector systematics and to optimize the detector's design.
This was performed by systematically varying a large number of detector
parameters and observing what consequence this had on the detectors ability to
carry out a good Partial Wave Analysis. Our results show, most noticably, that
even slight imperfections in the detector, can cause a strong signal in the
S-wave decay to feed into the corresponding D-wave decay. However,
at least with the cocktail of states used in this analysis, it was quite
dificult to get feed-through into the exotic, (
channel).
Monte Carlo Events
In this study the GENR8 program was used to produce a sample of final state
events. Approximately
million events of the following two types were
generated.
These events were then weighted with a physics hypothesis based on a
one-pion exchange model in which the resonances
and
would decay
via a
state to the
final state. The physics assumed that
the
, the
and the
were produced. The
was then allowed to deacy into
in both an S- and a D-wave.
The
was allowed to decay to
in both a P- and an F-wave.
The resulting events were then run through the Hall D detector Monte Carlo,
HDFAST. Approximately
events survived in each of the
two reaction channels.
Reaction 1 involved the detection and reconstruction of only
charged pions. This was used to test detector components designed to detect
charged particles. The changes tested with this reaction are:
- The Magnetc Field:
A mismatch was made between the magnetic field with which the physics events
were tracked, and between what was used in the normalization integral. While
this exact change is unlikely to affect the detector, it is easy to make
to the geometry and produced a good starting point to the entire study. Its
effect is to produces a global smearing of tracking parameters.
- The Forward Drift Chamber Resolution:
Earlier resolution studies indicated that the resolution of the the forward
drift chamber system was an important parameter. It is clear that this system
is responsible for detecting fast forward particles, and the best momentum
resolution is achieved with the best resolution in the forward chambers.
- The Forward Drift Chamber Beam Hole Size:
In this study a mismatch was created bewteen the size of the beamline hole in
forward chambers between the physics events samples, and the normalization
events. This attempts to simulate a problem in understanding the efficiency
in the very forward directions.
Reaction /refeq:neutral focused on the detection of the
's coming
from the two
decays. This reaction was used to test the detector
components specifically designed to detect photons, namely the Barrel
Calorimeter and the Lead Glass Detector. The changes tested with this reaction
are listed below.
- The low energy threshold of the Barrel Calorimeter.
Here we varied the minimum photon energy that could be reconstructed in the
Barrel Calorimeter. The nominal design value is
.
- The low energy threshold of the Lead Glass Detector.
Here we varied the minimum photon energy that could be detected in the
Pb-Glass system. The nominal value is
.
- Resolution Parameters for the Barrel calorimeter.
Two parameters control the resolution of the Barrel Calorimeter. A constant
offset and a percentage resolution parameter. Here we varied both of these
for the Barrel Calorimeter.
- Resolution Parameters for the Lead Glass Detector.
A similar set of parameters control the behavior of the Lead-Glass detector.
These parameters have been varied here.
Both reactions were simulated using a photon beam energy of
.
Each event was run through the HDFast detector simulation with the varied
geometry parameters, and then PWA was performed on the resulting data. The
data used to normalize the PWA was also run through the same detector
simulation. The physics channels simulated in the data are given above.
The goal of this study was to see the signal strength in a channel that
was not put in the original cocktail, leakage. The following set of partial
waves are physically possible to photoproduce in
reactions.
- A
particle,
. This is the very inetersting wave
as it has exotic, or non-
-
quantum numbers.
- An
particle,
, which would be produced in either
the
or
magnetic substate.
In addition to the above two, it is not possible to physically produce an
in either the
or
magnetic substates, but we could have leakage
into this state. Finally, a critical source of leakage would be to look at
one of the produced signals which is quite strong leaking into one of the
weaker channels. The two examples of this in this data set are the
decaying via S-wave leaking into the D-wave decay, and the
decaying
via P-wave leaking into the F-wave.
Results
PWA results for
reaction
The nominal results for reaction 1, (
)
are shown in Figures 1 and 2.
There is virtually no leakage into the any of the nonexistent channels as
seen in 2.
Figure:
PWA results for reaction 1
using the default geometry specifications. These results show the intensity
for everything, and then the three major components,
decaying via
S-wave, or
,
or
and the
decaying via
P-wave, or
. The bottom two figures show the weaker
D-wave
and the
F-wave results.
 |
Figure:
PWA results for reaction 1
using the default geometry specifications. These results show the intensity
for the waves that are not physically present in the data set. The
wave is the exotic
channel. The bottom two figures show the strength
of the leakage into the exotic
wave compared to the strength of the
signal -- the latter is believed to be the source of the leakage.
The level of the leakage is below 1% for the nomical detector configuration.
 |
Smeared Results for Reaction 1
A series of smearings were used to probe the effects on the partial wave
analysis.
Initially, the magnetic filed map used to track the physics
events was distorted with respect to that used in the normalizations.
Table 1 shows the various filed values tested under this
scenario. For the most part, these variations had little effect on the PWA
results. Figure 3 shows the feed-through into the exotic
channel for the case of
and
changes to the magnetic
fields. These are fairly extreme changes to the field strength, and not
typical of what would be expected under normal running conditions. These
results indicate that globally, the charged tracking system is good.
Table:
Magnetic field values used in the charge testing.
The nominal field value is
.
|
|
Figure:
Feed-through into the exotic
wave
when the field values have been distorted by
(left) and
(right).The signal is compaed to the strength of the
channel
which is belived to be the origin of this feed through. The intensity
of the feedthrough is about 1% of the
signal.
 |
A typical acceptance problem in partial wave analysis is that the exact
boundaries of a hole are not well understood. In order to study this, we
varied the size of the hole around the beamline in the forward detectors.
The nominal hole along the beamline is
radius. In performing these
studies, the hole size in the normalization integral was left alone, but that
in the physics events was varied. The range of hole sizes are shown in
Table 2, while the leakage into the exotic waves are shown
in Figure 4. As can be seen, the leakage into th exotic
wave is very small, even for extreme changes. Actually, the reason for this
is that to produce an exotic wave, we need to break a forward-backward or
up-down symmetry in the detector. This sort of change will produces no
such asymmetry. However, this sort of change can distort an even
distribution to look like a higher order even distribution. For
example, it would be posible to produce leakage from a strong S-wave into
a weaker D-wave. Such an effect can be seen in
Table:
Hole size in the forward direction used in
simulating the physics events.
|
|
Figure:
These plots show the exotic
wave for changes in the forward
hole geometry. The left plot is with the central hole radius of the FDC at
(default is 3.5cm). The right picture is with the whole size set
to
. Both signals are compared to the
signal, which is
believed to be the source of the leakage. Neither of these changes produces
significant leakage above 1% into the exotic wave.
 |
The nominal resolution of the forward drift chamber systems is
.
We have also run a conparison with this resolution degraded to
in reconstructing the physics events, but left at the nominal value for the
normalization events. The results of this are shown in Figure 5
Figure:
The left hand plot shows the leakage into the exotic
wave compared to
the
wave. The right hand figure shows the leakage into the
D-wave that comes from the
S-wave. Typical leakage into the exotic
wave of about 1% is observed, while a large leakage into the D-wave is seen.
 |
PWA results for
final state.
Reaction 2,
was simulated for
the primary purpose of testing the neutral particle detectors. The barrel
calorimeter and the lead glass detector. Both of these detectors have a
minimum detectable photon energy, which in the standard Monte Carlo is
for the Barrel Calorimeter and
for the Lead Glass
detector. In addition, there is a two parameter photon energy resolution
that can be varied for each detector. The results for standard geometry are
shown in Figures 6 and 7.
Figure:
These PWA plots are for reaction 2
(
) with default geometry specifications.
 |
Figure:
PWA results for reaction 2
using the default geometry specifications. These results show the
intensity for the waves that are not physically present in the data
set. The
wave is the exotic
channel. The bottom two
figures show the strength of the leakage into the exotic channel
compared to the
channel.
 |
Table:
The design value for the detector is
.
Results were also run using
and
as the low energy
cutoff for physics events as seen in the barrel calorimeter.
|
|
Figure:
These plots show the leakage in to the exotic
wave for a
minimum detectable photon energy of
and
in the barrel
calorimeter. The leakage is compared to the strength of the
signal, which is expected to be its source.
 |
Table:
The design value for the detector is
.
Results were also run using
and
as the low energy
cutoff for physics events as seen in the barrel calorimeter.
|
|
Figure:
These plots show the leakage in to the exotic
wave for a
minimum detectable photon energy of
and
in the
Lead Glass calorimeter. The leakage is compared to the strength of the
signal, which is expected to be its source.
 |
The parametrization of the barrel calorimeter showers are controlled by
two parameters as given in the following formula:
where the energy is expressed in units of
. The default values
for the two parameters are:
and
.
During these studies, we examined the following combinations as given in
Table 5. In looking at the results in Figure 10,
it at first appears that the increase in
as shown in the
middle figures leads to a clear feed through at the few percent level.
However, the plots for an even larger value of
do not
appear to confirm this effect. As such, it is difficult to conclude
that this is going to be a problem.
Table:
Tested resolution combinations for the the
Barrel Calorimeter.
|
|
Figure:
Changes to the Barrel Calorimeter resolution
parametrization.The left hand picture has
and
, the middle picture has
and
, finally the right hand picture has
and
.
 |
The parametrization of the lead glass calorimeter showers are controlled by
two parameters as given in the following formula:
where the energy is expressed in units of
. The default values
for the two parameters are:
and
.
During these studies, we examined the following combinations as given in
Table 6.
Table:
Tested resolution combinations for the the
Barrel Calorimeter.
|
|
The most important conclusions of this report is that it is difficult to
produce feed through into the exotic channel from other meson channels.
For almost all changes made here, the amount of feed through was less
than
of a strong channel, with the feed through for the nominal
design values being something like
.
However, we do see significant leakage from the
S-wave decay
into the
D-wave decay, with even small changes in the nominal
detector design. This sort of feed through is fairly straight forward
to understand. An S-wave decay is nominally flat, however, if we have
losses near
, the easiest description of this is with
a D-wave component. In order to produce a P-wave component, it is necessary
to produce a forward-backward asymmetry in the Jackson frame - something
that apepars fairly difficult to do with this detector.
Curtis A. Meyer
2001-12-07